Laminar flow cabinet with vertical air flow (Class I Biological Safety Cabinet )

Purpose: installed in the laboratory for pharmakinetic studies.

Design features:

  • table/support 1250x670x850 mm
Specifications

External overall dimensions DxWxH, mm:

  • 1250x670x1180

Working area dimensions DxWxHxH, mm:

  • 1150x600x670

Material of execution and equipment:

  • frame - steel AISI304 S-1,2mm
  • pipe - steel AISI304 (30x30x1,2 mm)
  • motors - EBM-PAPST - 220-380V
  • filter module class G4 - H14
  • diffmanometer MAGNEHELIC series 2000

Control unit:

  • SIEMENS LCD display (Logo)
  • automatic maintenance of air flow rate - 0.45 m/s, with the possibility of manual adjustment